• On The Insider: Bruno Film Edited Due to Jackson's Death

June 26, 2008 8:11 AM PDT

RFID safe for medical monitoring?

  • 3 comments
Radio frequency identification technology can interfere with medical equipment, according to a study published in The Journal of the American Medical Association.

The study by six Dutch scientists has found that electromagnetic interference, or EMI, generated by the tracking-and-tracing chips has the potential to cause disruption to medical devices, which could be a potential hazard to patients.

RFID-tagged wristbands are currently used in some hospitals to ensure that correct procedures are carried out on patients. But the future of health care is likely to involve far more RFID-style tech, according to a recent report by the U.K. Office of Communications, or Ofcom, which included a prediction that in-body and on-body wireless networks could become the norm over the next decade for monitoring patients.

The Dutch scientists looked at both active (info-transmitting) and passive (requiring a reader device) RFID systems, and they found that 34 adverse medical "incidents" were triggered out of 123 EMI tests--of which 22 were classed as hazardous, two as significant, and 10 as light, according to a critical-care scale. The average distance between the RFID reader and the medical device in all the EMI incidents was 30 centimeters (about a foot).

Passive RFID systems caused more incidents than active signals: 26 incidents in 41 EMI tests (63 percent), compared to eight incidents in 41 EMI tests (20 percent).

A report abstract states: "In a controlled nonclinical setting, RFID induced potentially hazardous incidents in medical devices. Implementation of RFID in the critical-care environment should require on-site EMI tests and updates of international standards."

The tests took place in May 2006 at the Academic Medical Center in the University of Amsterdam, using an active 125KHz RFID system and passive 868MHz in the proximity of 41 medical devices in 17 categories with 22 different manufacturers. Three tests were carried out on each piece of medical equipment.

Natasha Lomas of Silicon.com reported from London.

See more CNET content tagged:
RFID, EMI Group Plc., medical device, incident, medical equipment

Add a Comment (Log in or register) (3 Comments)
  • prev
  • 1
  • next
by franatwork June 26, 2008 10:15 AM PDT
RFID. Big Brother here we go.
Reply to this comment
by Eddie-c July 25, 2008 11:45 AM PDT
The study is incomplete - there are *4* bands certified for use and this study only uses 2. If a test is going to be done at least do it properly. ;)
Reply to this comment
by Philscbx August 25, 2008 10:31 PM PDT
What the heck, Why not just use simple BlueTooth to track patients.
They probably don't wonder outside of range.
If they do go out of range, that should alert central desk to check that person more than once a day maybe.
Reply to this comment
(3 Comments)
  • prev
  • 1
  • next
advertisement

Latest tech news headlines

RSS Feeds

Add headlines from CNET News to your homepage or feedreader.

More feeds available in our RSS feed index.

Markets

Market news, charts, SEC filings, and more

Related quotes

Dow Jones Industrials (-0.04%) -3.03 8,277.71
S&P 500 (-0.39%) -3.53 892.89
NASDAQ (-1.19%) -21.32 1,775.20
CNET TECH (66.33%) 510.74 1,280.75
  Symbol Lookup
advertisement

Inside CNET News

Scroll Left Scroll Right